11 Initial startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10 Service routines for startup
båÖäáëÜ
ZP: Adjusting and saving the zero point
1. Take the occlusal bite block out of the bite block holder.
"ZP" (for zero point) is displayed in selection field 1.
The Memory key lights up.
2. To save the zero point "ZP", first press the Memory key (the R key
lights up) and then the R key.
The value for the zero point "ZP" is accepted.
U: Adjusting and saving the upper value
1. Insert the occlusal bite block into the bite block holder and deflect the
bite block plate all the way up (up to the stop).
"U" (for upper value) is displayed in selection field 1.
The Memory key lights up.
2. To save the upper value "U", first press the Memory key (the R key
lights up) and then the R key.
The value for the upper value "U" is accepted.
L: Adjusting and saving the lower value
1. Deflect the bite block plate all the way down (up to the stop).
"L" (for lower value) is displayed in selection field 1.
The Memory key lights up.
2. To save the lower value "L", first press the Memory key (the R key
lights up) and then the R key.
The value for the lower value "L" is accepted.
The adjustment of the AD values is now finished.
Checking the adjustment
1. Check the adjustment via service routine S018.7 [ → 158]. To do this,
move the bite block plate and check whether the adjustment range is
between -17° and +22°.