11 Initial startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10 Service routines for startup
båÖäáëÜ
11.10.3.11
S017: Test step 15
* Factory setting
Activate/deactivate the acoustic signal for end of exposure
S017_5 GALILEOS / XG 3D / XG 3D ready
S017_5 XG 3D
1. Call service routine S017.15 [ → 125].
Once the service routine has been selected, the code for the
current setting is displayed in selection field 1.
2. Use the arrow keys (A) to select the required setting in selection field
1 (see table).
Once the required setting has been selected, the Memory key
lights up.
3. Save the setting [ → 129].
4. Exit the service routine [ → 129].
Selection field
Code
Function
1
00
Acoustic signal indicating the end of the
exposure is disabled
01
Acoustic signal indicating the end of the
exposure is enabled*