63 03 452 D3352
260
D3352
.
031.05.08
.
02 10.2015
13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.3 Checking and adjusting the touchscreen
Installation Manual ORTHOPHOS XG 3D / Ceph
6. Touch the center of the green square in the bottom left corner of the
screen.
The fifth adjustment screen appears on the display.
7. Confirm the adjustment settings by clicking the
"Yes"
button.
The touchscreen is now adjusted.
To repeat the adjustment procedure, touch the
"No"
button.
To abort the adjustment procedure, touch the
"Abort"
button.
Summary of Contents for Orthophos XG 3D/Ceph
Page 1: ...k W f j NMKOMNR loqelmelp ud Pa L b b Cover page...
Page 2: ......