63 03 452 D3352
196
D3352
.
031.05.08
.
02 10.2015
13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.1 General information about unit adjustment and calibration
Installation Manual ORTHOPHOS XG 3D / Ceph
13.1.2
"Adjustment/Calibration" menu
The menu guides you through the procedure to adjust and calibrate the
unit.
13.1.2.1
Calling the "Adjustment/Calibration" menu
Calling XG and GALILEOS, new
You can call the
"Adjustment/calibration"
menu via SIDEXIS XG:
"Utilities"
/
"Constancy test..."
/
"3D"
/
"Select X-ray device"
/
"Service
exposure "
/
"Select X-ray component"
/
"Adjustment/calibration"
/
Password prompt (see section entitled "Password protection")
The
"Select X-ray device"
and
"Select X-ray component"
prompts are
only displayed if more than one unit has been set up in SIDEXIS XG.
Password protection
Password XG and GALILEOS, new
The
"Adjustment/calibration"
menu is password-protected.
Reverse the number of the month and the number of the day and use
them as the password:
March 12th = 03 12 becomes 3021, which is the password of that day.
Service mode
Service XG and GALILEOS, new
When you open the
"Adjustment/calibration"
menu, the unit switches
from user mode to the PC service mode logged by the PC. In PC service
mode, the control options that are available on the control panel are
determined by SIDEXIS XG and the service routine currently selected.
General control of the unit by means of the control panel (as in the user
mode) is not possible in this mode.
Easypad service
Service mode is displayed on the Easypad via the PC service image.