13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2.1.3
Pan symmetry
13.2.1.3.1
Automatic adjustment: Pan symmetry
✔
The OP XG needle phantom is in the pan bite block holder on the unit.
1. In the structure tree, under
"2D Adjustment"
, click on the
"Pan -
Symmetrie"
element (S010.2).
The
"Pan - Symmetrie"
menu is displayed in the action area.
2. Make SIDEXIS XG ready for exposure. [ → 206]
3. Take an exposure (60 kV/3 mA; 14.1 s) [ → 206].
SIDEXIS XG calculates the adjustment values S1, S2, and S3
automatically from the exposure and enters them in the text
boxes of the
"Pan - Symmetrie"
menu.
The exposure is displayed in the exposure window in SIDEXIS
XG.