13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2.1.7
Ceph - Fixed point of rotation for QuickShot
13.2.1.7.1
Automatic adjustment: Ceph - QuickShot fixed point of rotation
(standard)
Adjustment 1
✔
The OP XG ceph adjustment phantom is
not
in the pan slot on the
unit.
1. In the structure tree, under
"2D Adjustment"
, click on the
"Ceph -
Fixed point of rotation Quickshot"
element (S010.8).
The
"Ceph - Fixed point of rotation Quickshot"
menu is displayed
in the action area.
2. Make SIDEXIS XG ready for exposure. [ → 206]
3. Take an exposure (80 kV/14 mA; 0.3 s) [ → 206].
Adjustment value S1 is calculated automatically by SIDEXIS XG
from the exposure and entered in the text box of the
"Ceph -
Fixed point of rotation Quickshot"
menu.
The exposure is displayed in the exposure window in SIDEXIS
XG.
NOTICE
Risk of damage to unit
It is essential that the OP XG needle phantom is removed from the bite
block holder on the device before a ceph exposure is taken; otherwise,
the combisensor could collide with the phantom.