63 03 452 D3352
D3352
.
031.05.08
.
02 10.2015
203
Sirona Dental Systems GmbH
13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.1 General information about unit adjustment and calibration
båÖäáëÜ
Pan - Diaphragm
Diaphragm adjustment can usually be performed directly via precision
adjustment. Only in exceptional cases, e.g. if the exposed image area is
completely outside the image section in an image acquired with the
"Precision adjustment"
setting (C), is it necessary to perform coarse
adjustment prior to precision adjustment (D).
On the exposure with coarse adjustment (D), the exposed area is just
barely visible at the right margin of the image field. Even in this extreme
case, an adjustment would still be possible.