11 Initial startup
Sirona Dental Systems GmbH
11.10 Service routines for startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10.4.7
S018: Test step 8 (occlusal bite block only)
Functional test of the occlusal bite block
All of the monitoring functions of the bite block electronics are active in
this service routine. Any error states are displayed as a code in selection
field 3.
1. Call service routine S018.8 [ → 125].
The current angle of the bite block plate or the one calculated
based on the calibration is displayed in selection field 1.
The currently measured AD value leading to the angle calculation
is displayed in selection field 2.
The codes of the possible error modes of the acquisition
electronics are displayed in selection field 3 (see Codes of
possible error modes [ → 161]).
The display of patient symbol key 1 (G) indicates whether or not
the bite block is connected (see table).
2. Exit the service routine [ → 129].
Symbol on the control
panel
Status
Meaning
Patient symbol key 1 (G) lit
Occlusal bite block
connected
not lit
Occlusal bite block not
connected
Selection field
Display
1
Angle of the bite block plate calculated on the basis of
the calibration
2
Measured AD value providing the basis for the angle
calculation
3