10 Installation: Ceph arm
Sirona Dental Systems GmbH
10.3 Installation
Installation Manual ORTHOPHOS XG 3D / Ceph
Positioning the ceph arm
1. Position the cardboard support next to the unit.
2. Position the ceph arm with the styrofoam packaging on top.
3. Remove the adhesive tape and the cover (E) from the ceph
packaging.
4. Break off the wall part (F) of the packaging at the perforation.
5. Slide the cardboard support with the ceph arm up to the wall so that
the drill holes of the support arm are horizontally aligned with the
positions of the threaded bolts.
6. Switch on the unit (G).
During the unit startup phase, the message
"System is starting"
appears on the control panel.
The unit's self-adjustment routine begins.
7. Acknowledge the error messages that appear using the R key (H) on
the control panel.
The main menu and help message H301 appear. You are
requested to move the unit to its starting position.
8. Press the R key (H) on the control panel.
The unit moves to its starting position.
CAUTION
X-rays
Be sure to observe the radiation protection regulations applicable in
your country.
➢ No person may be positioned in the unit when it is switched on.