13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
båÖäáëÜ
5. Repeat the procedure starting at Step 3 until the values in the menu's
text boxes are equal to "zero". This means that adjustment was
performed successfully.
IMPORTANT:If automatic adjustment fails to produce values equal to
zero, calculate the adjustment values manually and overwrite the
values in the menu's text boxes with these values [ → 214].
6. If adjustment is OK (the adjustment values in the text boxes are equal
to zero), save the values. [ → 206]
7. Remove the OP XG needle phantom from the pan bite block holder
on the unit.
8. Continue with the next stage of the adjustment procedure [ → 216].
A
Adjustment OK
● The three needle images must lie in the
center and inside the auxiliary lines.
B
Adjustment not OK