8 Electrical connection
Installation Manual ORTHOPHOS XG 3D / Ceph
8.1 Connecting the control cables
båÖäáëÜ
8
Electrical connection
8.1
Connecting the control cables
1. Use the media converter to connect the PC to the SC:SC socket on
the unit (see Operating Instructions of the media converter).
2. Install the media converter at a suitable location using fastening
screws or the Velcro strap supplied for this purpose. Do
not
attach the
media converter to the X-ray unit.
Installation version without remote control
➢ Connect the coiled cable of the release button to socket X103 of the
unit.
Installation version with remote control
1. Use cable L117 to connect the remote control to socket X103 on the
unit.
2. Secure the connector on socket X103.
A
PC
C
Media converter
B
Switch (not included in the
scope of supply)
D
Coiled cable
IMPORTANT
Use a duplex patch cable (L25), 1:1 connection, SC/SC 50/125 µm.
C
B
A
Wagoklemme
SC1:1
SIDEXIS
0
0 0 0 0 0
0 0
X103
D
X101
X108
L117
L25
Kat.5
RJ45
DX42 / X108
L108 / X103