11 Initial startup
Sirona Dental Systems GmbH
11.10 Service routines for startup
Installation Manual ORTHOPHOS XG 3D / Ceph
11.10.4.6
S018: Test step 7 (occlusal bite block only)
Setting the nominal angle for the occlusal bite block (as required by
the customer)
1. Call service routine S018.7 [ → 125].
The currently measured angle of the bite block plate is displayed
and permanently updated in selection field 2 (possible deflection:
-17° to +22°).
The display of patient symbol key 1 (G) indicates whether or not
the bite block is connected (see table).
Symbol on the control
panel
Status
Meaning
Patient symbol key 1 (G) lit
Occlusal bite block
connected
not lit
Occlusal bite block not
connected
Selection field
Parameter/Display
Range of values
1
Nominal angle of the bite block
plate
-17° to +22°
2
Measured deflection of the
bite block plate
-17° to +22°