13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.1 General information about unit adjustment and calibration
Installation Manual ORTHOPHOS XG 3D / Ceph
Validity of existing adjustment/calibration settings
The elements of the structure tree used for adjusting (calibrating) the
device are prefixed by symbols indicating the current status of the
corresponding adjustment or calibration operation.
The elements for resetting adjustments (calibrations) are not prefixed by
symbols.
2D traffic light adjustment
2D adjustment
3D traffic light adjustment
3D adjustment/calibration
Working with the structure tree
You can navigate between the elements of the structure tree by clicking
on them with the mouse.
NOTICE! You must follow the prescribed sequence in order to obtain a
valid adjustment or calibration.
Click with the mouse on the small triangles in front of the elements to
collapse and expand the structure tree.
Not checked Invalid data record or no
record present
Adjustment necessary
Checked
Valid data record,
adjustment performed and
saved (up to date)
No adjustment necessary
Green and
checked
Valid data record;
adjustment/calibration is in
progress
No adjustment/calibration
required
yellow
Data record available,
but not yet
saved
Data record must be saved
Red
Invalid data record or no
record present
Adjustment/calibration
required