13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2.1.1.2
Manual adjustment: Panoramic sensor
The manual adjustment procedure is similar to the one for automatic
adjustment [ → 212]. The only difference is that the default adjustment
values calculated automatically by SIDEXIS XG in the
"Pan - Sensor
Adjustment"
menu are overwritten by adjustment values calculated
manually.
1. Start the adjustment procedure as described in the chapter entitled
"Automatic adjustment: Panoramic sensor (standard) [ → 212]".
2. After taking the exposure, measure the distances S1, S2 and S3 with
the SIDEXIS XG measuring ruler.
To measure S1, S2 and S3, estimate where the horizontal center of
the displayed needles is located. Take the measurement in the lower
area of the needles if possible, since they may have become slightly
bent after repeated use.
Tip: To facilitate the measuring procedure, you can also color the
image in SIDEXIS XG (see the SIDEXIS Operator's Manual).