13 Adjusting and calibrating the unit
Installation Manual ORTHOPHOS XG 3D / Ceph
13.2 Adjustment and calibration via the "Unit adjustment/calibration" menu
båÖäáëÜ
3. Overwrite the adjustment values for S1, S2 and S3 in the text boxes
of the
"Pan - Sensor Adjustment"
menu with the measured values.
For information about the shifting direction (input of +/- sign in the
menu), see the chapter entitled "Shifting direction of the exposed
image area". Use points as decimal separators!
4. Continue with the adjustment procedure as described in the chapter