13 Adjusting and calibrating the unit
Sirona Dental Systems GmbH
13.1 General information about unit adjustment and calibration
Installation Manual ORTHOPHOS XG 3D / Ceph
13.1.3
Enabling exposure readiness
To take an exposure in SIDEXIS XG, the system must first be made ready
for exposure.
XG and GALILEOS ready for exposure, new
✔
Call the
"Adjustment/calibration"
menu.
✔
Select the corresponding element in the structure tree.
➢ Click on the
"Image acquisition"
button.
The exposure window opens in SIDEXIS XG. It indicates the
current status of exposure readiness.
The service routine used for the corresponding exposure is
displayed on the control panel, along with the specific exposure
parameters.
13.1.4
Taking an exposure
Taking an exposure with XG 3D and GALILEOS, new
✔
Call the
"Adjustment/calibration"
menu.
✔
Select the corresponding element in the structure tree.
✔
SIDEXIS XG must be ready for exposure.
➢ Press the release button. Press and hold down the button until the
exposure is complete, the preview image is displayed in the exposure
window, and the acoustic signal that indicates the end of the
exposure (double beep) sounds (if it has been configured).
13.1.5
Save values
Saving values XG3D
✔
Adjustment or calibration must be OK, i.e. the values in the menu's
text boxes must be equal to zero.
➢ To save the adjustment or calibration values, click the
"Save values"
button.
The adjustment or calibration is saved.
The saved adjustment or calibration is identified in the structure
tree by a check mark (2D adjustment) or a green traffic light
symbol with a check mark (3D adjustment).