Rapidlab 1200 Operator’s Guide: Troubleshooting
6-63
02087462 Rev. V
D23 Reagent Flow Error: 14
D24 Codes
D24 AQC Material Error: 1, 2, 3
D32 Codes
D32 AQC Cartridge Valve Error: 1
Problem
Corrective Action
The system detects
a possible
obstruction during
calibration or
during the wash
cycle after sample
analysis.
1. Select
Diagnostics > Fluidic Functions > RCx
and look for
obstructions.
2. Remove obstructions from reagent manifold, measurement
module, and sample tubing.
3. Resolve any other D codes.
4. Perform the RCx test again.
5. Select
Diagnostics > Fluidic Functions > Wash Cycle
.
6. If the problem persists, replace the wash cartridge.
Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
7. If the problem persists, replace the reagent cartridge.
Perform the highlighted test(s) in
Diagnostics > Fluidic
Functions
.
8. If the problem persists, call for technical assistance.
Problem
Corrective Action
An error was
detected during
delivery of AQC
material.
1 level 1
2 level 2
3 level 3
1. Check that the AutomaticQC connector is completely closed
and that the sensors are aligned correctly.
2. Select
Diagnostics > Fluidic Functions > Prime AQC
Cartridge
.
The correct button is highlighted.
3. Check if there is an obstruction in the reagent probe
cartridge.
Refer to
Flushing the Tubing for the AutomaticQC Module
4. If the problem persists, replace the AutomaticQC cartridge.
NOTE:
If you are not able to replace the cartridge, turn off the
AutomaticQC option in Setup to continue analyzing patient
samples.
5. If the problem persists, call for technical assistance.
Problem
Corrective Action
A valve does not
move to the correct
position.
1. Select
Diagnostics > Pumps and Valves > AQC Valve
.
2. If the problem persists, replace the AutomaticQC cartridge.
3. If the problem persists, call for technical assistance.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...