Rapidlab 1200 Operator’s Guide: System Configuration
8-3
02087462 Rev. V
Entering Customized QC Range Limits
1. At the Setup screen, select
QC
.
2. If performing AQC, select
AutomaticQC Ranges
.
If performing RQC, select
Required QC Ranges
.
3. Select the QC range to view.
The screen for the QC range you selected displays.
4. Select a parameter.
5. Enter a low limit, a high limit, or both, in the Low and High boxes, for the
selected parameter.
If in RiliBÄK mode, use the following equations to determine low and high
values:
•
Range Offset from Manufacturing Target =
Manufacturing (Manufacturing Target * Desired Deviation from
Target)
•
QC Low Value =
(Manufacturing Target – Range Offset from Manufacturing Target)
•
QC High Value =
(Manufacturing Range Offset from Manufacturing Target)
6. Repeat steps 4-5 as needed, for additional parameters.
7. Select
Continue
.
Setting Up Required QC
Use the Required QC screen to set up a required QC sample schedule. Use the
barcode scanner and RapidQC controls to perform Required QC analysis.
NOTE:
Ensure that you enter information for each lot of the controls you
schedule. Refer to
‚ page 8-4. You cannot analyze
the scheduled controls until you enter the lot information.
1. At the Setup screen, select
QC > Required QC Schedule
.
The screen displays the Required QC Schedule with the current schedule for
Required QC analysis, if defined.
NOTE:
When you change intervals, the system clears any times and levels
that were previously scheduled.
2. Select
Intervals
.
The system displays the Intervals screen.
3. Select the number of hours for the interval of the Required QC analysis.
For example, select 8 to have an 8-hour window for performing scheduled
controls.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...