
Rapidlab 1200 Operator’s Guide: Maintenance
5-37
02087462 Rev. V
b. Stand the internal electrode on its cap to drain it.
3. Fill the internal electrode compartment of the new sensor with KCl fill
solution:
a. Insert the tip of the KCl fill solution container into the internal electrode
compartment.
Figure 5-19 Reference Sensor Cassette and KCl Fill Solution
b. Fill the internal electrode compartment until the KCl fill solution enters
the KCl reservoir.
4. Install the internal electrode into the internal electrode compartment and
screw the electrode into place using the hex tool.
1
KCl fill solution
2
KCl reservoir
3
Internal electrode compartment
CAUTION:
Do not touch the internal electrode wire. The wire is fragile and is
easily damaged.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...