Rapidlab 1200 Operator’s Guide: Operating the System
2-13
02087462 Rev. V
NOTE:
If you have a priority sample, but the system displays a message that
the system is busy, select
Cancel
to interrupt the system. If Cancel is not
available, wait until the system is not displaying the busy message to analyze
the patient sample.
2. At the Analysis screen, select the button for capillary sample type.
3. Scan the barcode for the patient sample, if required.
4. Hold the capillary tube at the end closest to the sample port and insert the tube
firmly in the sample port.
Figure 2-3 Rapidlab 1200 Systems Sample Port
5. Select
Analyze
.
NOTE:
If the system determines that insufficient sample is present or bubbles
are in the sample, you are prompted to position the sample as you would a
microsample. For information about analyzing microsamples, refer to
6. When prompted, remove the sample device and select
Continue
.
7. If prompted, enter demographic information and select
Continue
.
1
Inserting the capillary tube
CAUTION:
Do not leave the sample device in the sample port after the system
prompts you to remove it. The system performs a wash after each sample analysis.
Leaving the sample device in place while the system performs a wash could
contaminate that wash with blood and adversely affect the next sample or system
operation.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...