Rapidlab 1200 Operator’s Guide: Maintenance
5-61
02087462 Rev. V
Scheduling Maintenance Activities
The maintenance schedule displays the maintenance tasks that are currently
scheduled. The schedule displays a calendar week with the day and date at the top
of the schedule and the state of each task in the grid.
You can navigate from week to week by using the right and left pointing arrows at
the top of the right panel. You can navigate within the schedule grid, moving
down the task list or back and forth through the days of the week by using the four
directional arrows at the middle of the right panel. The currently selected cell has
a white background.
Performing Scheduled Maintenance Tasks
After performing a maintenance task, you must mark it Complete on the schedule.
Some tasks, like Deproteinizing the Sample Path, automatically mark the schedule
as Complete but you should check for the Complete mark after completing each
Maintenance task.
If you complete the task within the grace period, the task is automatically
scheduled for its next due date.
•
If you complete a task before the grace period, the system does not change the
schedule and you are prompted to perform that task again during its regularly
scheduled time.
•
If you complete a task after the grace period, the next task is scheduled for the
scheduled time period after you actually perform the task.
To record a maintenance task, perform the following steps:
1. Select
Status > Maintenance > Schedule
.
2. Select the maintenance task you want to record as completed.
3. Select
Completed
.
Task state
Background
Indicator
Overdue
red
[blank]
Due
yellow
solid circle
Not Due
green
[blank]
As Required
green
[blank]
Completed
checkmark
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...