5-38
Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
5. Fill the KCl reservoir with KCl fill solution:
a. Remove the reservoir cap using the hex tool and set the cap aside.
b. Partially fill the KCl reservoir.
c. Tap the front face of the sensor with your knuckle to remove any bubbles.
d. Continue filling the reservoir with KCl fill solution to the fill line.
6. Hand tighten the reservoir cap.
7. Clean and inspect the sensor:
a. Tap the front face of the sensor with your knuckle to remove any bubbles.
b. Wipe any excess fill solution from the exterior of the sensor using a lint-free
tissue.
c. Ensure that the O-rings are in place.
8. Install the Reference sensor.
Refer to
9. Verify sensor performance.
Refer to
CAUTION:
Do not overtighten the reservoir cap. Overtightening can deform the
gasket and cause leaks.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...