5-8
Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
4. Check the level of fill solution in the reference sensor:
a. Verify that the reference sensor fill solution is above the fill line.
The fill line is on the right side of the KCl reservoir.
NOTE:
The fill solution can become slightly pink as a result of analyzing QC and
AutomaticQC materials. A pink color that becomes very dark over a 12 hour
period indicates a pending failure of the reference sensor. When this occurs,
replace the reference sensor or the reference sensor cassette.
b. If the level of fill solution is low, fill the sensor.
Refer to
Performing Reference Sensor Maintenance
5. Check the level of fill solution in the K
+
, Cl¯, Ca
++
, Na
+
, and pH sensors:
a. Verify that the K
+
, Cl¯, and Ca
++
sensors are almost full with a small (1 mm)
bubble at the top.
b. Verify that the pH and Na
+
sensors are completely full.
c. If the level of fill solution is low, replace the fill solution.
Refer to
Performing Measurement Sensor Maintenance
6. Close the measurement module door and close the front cover.
7. Allow the system to warm up for 10 minutes.
8. Select
Yes
.
9. Select
Continue > Return
.
10. Select
Analyze > Yes
.
11. Analyze a minimum of 2 levels of quality control material to verify sensor
performance.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...