Rapidlab 1200 Operator’s Guide: Maintenance
5-13
02087462 Rev. V
b. Connect the new tubing to the pressure sensor bubbler and to the reagent
manifold.
6. Dispose of the tubing according to your institution’s protocol.
7. Close the reagent cartridge door and then close the front cover.
8. If you have a Rapidlab 1245 or 1265 system and are replacing the CO-ox
tubing, continue with step 2 of
Replacing the CO-ox Module Tubing
9. Close the measurement module door.
10. Select
No
.
11. Select
Return > Analyze > Yes
.
Replacing the CO-ox Module Tubing
You need a CO-ox tubing kit for this task.
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to
Appendix A, Protecting Yourself from Biohazards
for recommended
precautions when working with biohazardous materials.
1. Select
Status > Diagnostics > Cartridges
.
2. Select
Eject W Cartridge
and remove the wash cartridge.
3. Lift the door to the measurement module.
Refer to
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...