5-14
Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
Replacing the CO-ox Sample Tubing
1. Disconnect the sample tubing from the connector on top of the sample chamber.
Figure 5-4 Measurement Module Sample Chamber
2. Rotate the gray wheel towards you, pull the tubing through the fluid detector, and
remove the tubing from the steel measurement module connector.
3. Push the new tubing through the fluid detector.
4. Connect the tubing to the steel measurement module connector and to the sample
chamber connector.
Use a clamp or tweezers to attach the tubing.
1
Sample chamber connector
2
Fluid detector
3
Measurement module connector