
5-18
Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
d. Squeeze the finger snaps on the reagent manifold and then remove the manifold
by sliding it down and away from the sample connector.
Figure 5-8 Reagent Manifold Finger Snaps
e. Dispose of the manifold according to your institution’s protocol.
4. Install the new reagent manifold:
a. Snap the new reagent manifold in place.
NOTE:
When you are installing the tubing, avoid pinching or bending the tubing.
b. Connect the waste tubing.
c. For Rapidlab 1245 and 1265 systems, connect the reagent manifold tubing to the
CO-ox pump tubing at the fitting.
Completing Reagent Manifold Maintenance
1. Close the front cover.
2. Reinstall the reagent and wash cartridges.
3. Select
Return
.
4. Verify system performance:
a. Perform a 2-point calibration.
Refer to
Performing Manual Calibrations
b. Analyze a minimum of 2 levels of quality control material.
1
Finger snaps
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...