8-4
Rapidlab 1200 Operator’s Guide: System Configuration
02087462 Rev. V
4. Select
Save
.
The Intervals screen closes.
5. Schedule levels for each time and day, as necessary:
a. Use the arrow keys to navigate to a place in the grid indicating the day of the week
and time period for the level or levels.
b. Select one or more buttons for the level or levels to schedule.
A check mark on the button means that you scheduled that level. Select the button
again to remove the check mark.
Setting up Required QC Ranges
Use this screen to enter the following information for the controls you use for Required
QC analysis:
•
Control name
•
Lot number
•
Expiration date
•
Target ranges for each level
Required QC ranges support both an active and trial lot. Only the active lot is used to
satisfy the Required QC schedule. The trial lot is used to evaluate new materials.
To define the schedule your system uses for automatically analyzing controls, refer to
Setting up AutomaticQC Schedule
1. At the Setup screen, select
QC > Required QC Ranges
.
NOTE:
When you scan a lot for a level, it becomes the active lot. If you already have
an active lot for a level and scan in new lot information for the same level, it becomes
the trial lot.
When you scan the barcode, the system makes the button corresponding to that level
available. If a level button is available, that barcode is already scanned.
2. Scan the
catalog number barcode
on the QC Expected Values chart.
When you scan the barcode for a new lot of controls, the system displays a message
prompting you to keep the existing target ranges.
3. Scan the
lot number barcode
for a level of the control on the QC Expected Values
chart.
This displays the Range Entry screen. This screen displays the control name, level, lot
number, and expiration date under the banner portion of the screen. It also displays the
parameters and their ranges and allows you to edit the ranges. Refer to
4. At the Range Entry screen, scan the barcode for each parameter to enter the target
ranges for this level.
5. Select
Save
.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...