Rapidlab 1200 Operator’s Guide: Troubleshooting
6-31
02087462 Rev. V
Removing Obstructions from the CO-ox Sample Path
Use the following procedure to remove obstructions from the CO-ox sample path
where you see fluid restrictions and obstructions, or suspect problems based on
D code messages.
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to
Appendix A, Protecting Yourself from Biohazards
for recommended
precautions when working with biohazardous materials.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
1. Remove the reagent cartridge and the wash cartridge:
a. Select
Status > Diagnostics > Cartridges > Eject Both Cartridges
.
b. Remove both the cartridges.
c. Leave the doors open.
2. If the obstruction is between the sample chamber and the measurement
module connector, replace the CO-ox sample tubing:
a. Disconnect the sample tubing from the connector on top of the sample
chamber.
Figure 6-15 CO-ox Sample Path
1
Sample chamber connector
2
Fluid detector
3
Measurement module connector
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...