6-46
Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
Performing the Wavelength Cal Test
The Wavelength Cal test verifies the proper operation of the CO-ox neon lamp and prints
the results providing the system ID, the date and time of the report, and as represented in
Table 6-7.
To perform the Wavelength Cal Report, select
CO-ox > Wavelength Cal
.
Performing the CO-ox Sample Chamber Test
This test is only performed on Rapidlab 1245 and 1265 Systems.
Use this procedure to test the ability of the CO-ox sample chamber to move to the open
and close positions. The systems displays the pass or fail results on a printed report.
1. Select
Status > Diagnostics
.
2. Enter your password, if necessary.
3. Select
CO-ox > Start
.
The sample chamber moves to the open position.
4. Select
Change
to move the sample chamber to the closed position.
5. Select
Stop
.
The system prints the test results Passed or Failed.
6. Wait while the system returns the sample chamber to the correct position.
7. Select
Return
.
8. To return to the Status screen, select
Return
.
Table 6-7 Wavelength Cal Report
Row number
Data
1
Mean
2
Standard deviation
3
Measured position
4
Baseline position
5
Shift
6
Integration time from last zero
7
Serial number
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...