Rapidlab 1200 Operator’s Guide: Maintenance
5-59
02087462 Rev. V
Flushing the CO-ox Pump Tubing
This section applies to the Rapidlab 1245 and 1265 systems.
1. Remove the CO-ox roller cage.
2. Disconnect the CO-ox sample tubing from the connector on the measurement
module.
Figure 5-33 CO-ox Sample Chamber Tubing
3. Flush the tubing with a minimum of 3 mL water.
4. Flush the tubing with a minimum of 12 mL air.
5. Collect the water in an absorbent swab or equivalent.
6. Reinstall the CO-ox roller cage.
7. Reconnect the CO-ox pump tubing and the CO-ox sample tubing.
8. Remove the CO-ox sample chamber and discard it according to your
institution’s protocol.
9. To exit Diagnostics, close the cartridge doors.
10. Select
Return
.
1
Sample chamber connector
2
Fluid detector
3
Measurement module connector, collect the water here
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...