Rapidlab 1200 Operator’s Guide: Troubleshooting
6-15
02087462 Rev. V
8. Discard the gauze.
Figure 6-2 Location of Reagent Manifold Ports and Exit Ports
9. Repeat the procedure on the sample port.
a. Fill the plastic syringe with water, and place it against the sample port
b. Press the syringe plunger gently to insert water into the CO-ox waste port,
while holding gauze against the end of the tubing that is connected
to the CO-ox waste exit port (item 2, Figure 6-2), to catch the obstruction
and any excess liquid.
c. Discard the gauze.
1
CO-ox waste exit port (this illustration shows ports only and does not show
the reagent manifold tubing that remains connected to this port when you
perform this procedure)
2
Sample exit port
3
Sensor waste exit port
4
Sensor waste port
5
Sample port
6
CO-ox waste port
CAUTION:
Do not disconnect the reagent manifold tubing from the port on the
top, left-side of the reagent manifold.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...