Rapidlab 1200 Operator’s Guide: Maintenance
5-23
02087462 Rev. V
b. Push the spring-loaded latch to the right, grasp the tab on the reference
sensor, and pull the reference sensor up and out.
Figure 5-11 Measurement Module
c. Align the top of the test/blank reference sensor (TB5) with the sensor
contact and snap the sensor into place.
4. If you have a Rapidlab 1260 or 1265 system, replace the Glucose and Lactate
biosensors with the test/blank Glucose Lactate (TB4) sensors.
Refer to step 3.
5. Close the measurement module door and then close the front cover.
6. Aspirate the deproteinizer:
a. Invert the deproteinizer vial several times to mix the deproteinizer
solution.
b. Draw the deproteinizer into a syringe and insert the syringe into the
sample port.
c. Select
Continue
.
d. When prompted, remove the syringe and select
Continue
.
The system displays a message prompting you to condition the sensors.
1
Reference sensor
2
Spring-loaded latch
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...