Rapidlab 1200 Operator’s Guide: Maintenance
5-49
02087462 Rev. V
Relocating the System
Use this procedure to move the system for use in another area.
CAUTION:
Do not remove power from the system for more than 6 hours if
cartridges are installed. Cartridges installed in the system remain stable for 6
hours without power.
1. Select
Status > Shutdown
.
2. When the system displays the Shutdown screen, turn the power off and
disconnect the power cord from the electrical outlet.
3. Disinfect the exterior surfaces of the system.
Refer to
Cleaning and Disinfecting the Exterior Surfaces
4. Move the system to its new location.
5. To restore power to the system, connect the power cord to the electrical outlet.
6. Turn the power on.
WARNING:
Do not proceed while the power is on. To prevent electrical shock or
damage to the system, remove power from the system as described in this
procedure.
WARNING:
Do not operate the system in the presence of flammable anesthetic
mixture with air, O
2
, or nitrous oxide. The risk of explosion exists in a potentially
explosive environment. Refer to
Protecting Yourself from Electrical Hazards
page A-3 for recommended precautions when working around electricity.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...