Rapidlab 1200 Operator’s Guide: Operating the System
2-25
02087462 Rev. V
System Behavior During Microsample Analysis
•
If you finish entering data but aspiration is not complete, the Wait screen
displays. When aspiration completes you are prompted to remove the sample.
•
If you do not finish entering data, and aspiration completes, the system may
prompt you to remove the sample.
•
You may be prompted to verify the sample after the sample has been
positioned, if
p
O
2
and
p
CO
2
measurements are performed.
•
When data entry is interrupted any data not saved or partially entered is lost.
•
After an interruption to position or verify a microsample the following
behavior occurs:
•
You are returned to the Data Entry screen.
•
The field that was active when the interruption occurred is highlighted
when you are returned to the screen.
•
If you were at a submenu of the Data Entry screen when data entry was
interrupted, such as the alphanumeric keypad, you are returned to the
Data Entry screen, and the last active field in the Data Entry screen is
highlighted.
Using the Save Demographics Option
Use the Save Demographics option when you are analyzing samples for the same
patient. This option allows you to re-enter the demographic information from the
previous sample when analyzing a sample from the same patient.
To use this option during patient analysis, perform the following steps:
1. At the Data Entry screen, select
Last Patient
.
This automatically re-enters the demographics from the previous sample.
2. If the demographics from the previous patient are not correct for the current
sample, select
Clear All
.
This deletes the information. For more information about the Save
Demographics option, refer to
Selecting the Save Demographics Option
Editing Demographics
NOTE:
You can edit demographics if you turned on this option in Setup. For more
information about selecting this option, refer to
Selecting Demographics Editing
Use this procedure to edit patient or sample demographic data, such as the patient
name or temperature, for previously analyzed patient samples.
If you edit the name, sex, or date of birth, the system applies the changes to all
samples saved for the patient. You can also print a patient sample report with the
edited data and send the data to a data management system or an LIS.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...