Rapidlab 1200 Operator’s Guide: Troubleshooting
6-5
02087462 Rev. V
•
Check the solution level in the affected sensor and remove bubbles.
The K
+
, Ca
++
, and Cl¯ sensors should be nearly full. The pH and Na
+
sensors
should be completely full without any bubbles.
•
Check the sensor contacts and measurement block for salt buildup.
•
Check behind the sensors for fluid leaks that can come from the reference
sensor.
•
Check that the sensors are installed in the correct order and are aligned, the
O-rings are in place, and the spring-loaded latch is closed.
•
If the pH or Na
+
sensor have a D2 or D3, condition the sensors.
•
Check for cracks or leaks in the sample tubing, and replace the tubing if
necessary.
•
Replace the affected sensor, if necessary.
Using the Calibration Report to Identify Problems
The full calibration report shows the measurement and drift values. Rapidlab 1200
systems use a 2-point calibration curve. The two points on the curve are the Cal
Point and Slope Point.
•
The values in the Meas column represent the actual measured value of the
current calibration.
•
The values in the Drift column represent the difference between the measured
value from the current calibration and the value assigned to the analyte during
the manufacture of the cartridge.
•
If the system cannot report a value during a calibration, the result is blank on
the report and the D code and explanation prints at the bottom of the report
The following symbols can display with the results on the printed report.
Table 6-1 Symbols
↑
or
↓
This symbol...
Indicates that...
↑
or
↓
the result is above or below the drift limit.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...