5-36
Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
Filling the Reference Sensor Cassette
If you do not have a new reference sensor cassette, refill the reference sensor instead.
You need the following materials for this task:
•
Reference electrode refill kit
•
Reference electrode internal replacement kit
•
Hex tool
•
Lint-free tissue
1. Remove the sensor.
Refer to
2. Remove the internal reference electrode:
a. Remove the internal reference electrode using the hex tool.
Figure 5-18 Reference Sensor Cassette
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to
Appendix A, Protecting Yourself from Biohazards
for recommended
precautions when working with biohazardous materials.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
1
Internal electrode
2
Cassette
3
Internal electrode compartment
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...