Rapidlab 1200 Operator’s Guide: Troubleshooting
6-27
02087462 Rev. V
b. If you detect an obstruction, fill a syringe with water, place the syringe
against the sample path and inject water through the sample path.
Figure 6-11 Injecting Water into the Gas Sensor Sample Path
c. For the sample ground/temperature sensor, pH sensor, and reference
sensor, push the clot-removal line through the sample path to remove
obstructions.
Figure 6-12 Clot-Removal Line in Sensor Sample Path
1
Sensor
2
Syringe
CAUTION:
Use the clot-removal line carefully to avoid damage to the sample
ground/temperature, pH, and reference sensors.
1
Sensor cartridge
2
Clot-removal line