5-48
Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
1. Select
Status > Shutdown > Yes
.
The system displays the Shutdown screen and stops all fluidic sequences so that no
damage to the cartridges can occur.
2. Turn the power off and disconnect the power cord from the electrical outlet.
3. Open the fuse holder:
a. Insert a small flat-head screwdriver in the tabs on each side of the fuse holder.
b. Pull the fuse holder out from the compartment as far as possible.
Figure 5-24 Fuse Holder
4. Remove the old fuses and dispose of them according to your institution’s protocol.
5. Install new fuses.
6. Slide the fuse holder into the fuse compartment.
7. Restore power:
a. Reconnect the power cord to the system.
b. Reconnect the power cord to the electrical outlet.
c. Turn the power on.
1
Fuse holder
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...