6-82
Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
R Cartridge
Not Valid
The system detects that an
expired reagent cartridge was
installed or the cartridge was
installed after being removed for
more than 6 hours.
The system is unable to use the
cartridge.
Replace the cartridge. Refer to
Replacing the Reagent
Cartridge
Sample Temp
Out of Range
The system detects that the
temperature of the patient or QC
sample is beyond the acceptable
measurement range at the end of
sample analysis.
The system reports the sample
results. The message displays on
the printed report.
Check that the front cover is
closed tightly.
Analyze the sample again.
Scheduled QC
Not Done
(Level 1, 2, or
3)
AutomaticQC or Required QC
was not performed when
scheduled.
The system performs
AutomaticQC analysis for
missed QC when the system
becomes available.
Perform Required QC analysis,
or restore parameters.
SulfHb > 1.5% The system detects a
sulfhemoglobin concentration
greater than 1.5% in the sample.
The message displays on the
printed report.
Temp Not
Ready
The temperature of the
measurement module is beyond
the acceptable measurement
range.
This message can occur when the
system is warming up after being
shut down or when the door
remains open for too long.
Ensure that you closed the front
cover securely.
Wait until the system removes
the message from the events log
before analyzing samples.
Temp Warning
The system detects that the
temperature of the measurement
module is beyond the specified
range for sample analysis.
This message can occur when the
system is warming up after being
shut down or when the door
remains open for too long.
Ensure that you closed the front
cover securely.
Wait until the system removes
the message from the events log
before analyzing samples.
Message
Probable Cause
Corrective Action
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...