5-34
Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
Performing Reference Sensor Maintenance
The Reference sensor requires regular cleaning to maintain performance.
Cleaning and Inspecting the Reference Sensor
You need the following materials for this task:
•
Water
•
Lint-free swab
1. Remove the Reference sensor.
Refer to
Figure 5-16 Measurement Module
BIOHAZARD:
Wear personal protective equipment. Use universal precautions.
Refer to
Appendix A, Protecting Yourself from Biohazards
for recommended
precautions when working with biohazardous materials.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the
measurement module frame to discharge static buildup.
1
Reference sensor
2
Spring-loaded latch
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...