Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
D51 Lactate Sensor Error
D60 Communications Error: 1, 2
D70 Codes
D70 Optics Error: 2
D70 Optics Error: 3, 4, 7
Problem
Corrective Action
The system detects
an open connection
in the lactate
biosensor.
1. Check that the biosensor is correctly installed, the contacts
are aligned, and the spring-loaded latch is closed.
2. If the biosensor has been on the system less than one day,
perform three 2-point calibrations.
Refer to
Performing Manual Calibrations
3. If the problem persists, replace the biosensor.
4. If the problem persists, call for technical assistance.
Problem
Corrective Action
The system
detects a data
communication
error when trying to
communicate with
the Rapidlink or
Rapidcomm data
management
system or the LIS.
•
Ensure that the cable is firmly connected to the
Rapidlab 1200 system and to all external devices.
•
Ensure that the cable is not damaged and that it is the correct
cable for connecting the systems.
•
Inspect the cables for wear or crimps.
•
Ensure that the Rapidlink or Rapidcomm data management
system or the LIS is correctly configured to communicate
with the Rapidlab 1200 system.
•
Ensure that the Rapidlab 1200 system is correctly
configured to communicate with the Rapidlink or
Rapidcomm system or the LIS and that the connection is
turned on.
•
Select
Status > Shutdown
.
•
Wait 10 seconds, and then turn the system on.
•
If the problem persists, call for technical assistance.
Problem
Corrective Action
The system fails a
CO-ox optical
clarity test.
1. Perform a 1-point calibration.
Performing Manual Calibrations
2. If the problem persists, replace the CO-ox sample chamber.
3. If the problem persists, call for technical assistance.
Problem
Corrective Action
The system detects
a failure in the
CO-ox optical
components.
•
Perform a 2-point calibration.
Refer to
Performing Manual Calibrations
•
Select
Diagnostics > CO-ox > Lamp Cal
.
•
If the problem persists, call for technical assistance.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...