Rapidlab 1200 Operator’s Guide: Maintenance
02087462 Rev. V
4. Remove the waste assembly:
a. Lift the orange lever up.
Figure 5-14 Waste Assembly
b. Slide the waste assembly to the right.
5. Clean the waste assembly with the 10% bleach solution.
6. Replace the waste assembly and push the orange lever down.
7. Place the new waste bottle in the waste assembly.
8. Install a new AutomaticQC cartridge.
9. Select
Return
.
10. Dispose of the bleach solution according to your institution’s protocol.