6-78
Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
Excessive
Bubbles in
CO-ox Sample
The system detects bubbles in the
CO-ox sample chamber.
This occurs if the sample
contained bubbles, if air leaks are
present in the CO-ox sample
tubing, or if you inserted the
jagged end of capillary tubes into
the sample port.
CO-ox results are not reported.
Results for other enabled
parameters are reported.
1. Ensure that you use the
recommended storage,
handling, and mixing
techniques.
Refer to
page 2-7. For QC samples,
use the storage and handling
techniques recommended by
the manufacturer.
2. Analyze the sample again,
ensuring that the sample has
no bubbles before
introducing it into the
sample port. Checking for
bubbles is especially
important for samples
introduced from capillary
tubes.
3. Check the CO-ox sample
tubing for cracks or leaks.
4. When analyzing a capillary
sample, ensure that you
introduce the smooth end of
the capillary tube into the
sample port.
Insufficient
Sample
The system detects that not
enough sample volume exists to
complete routine analysis.
This is caused when you remove
the sample device from the
sample port before aspiration
completes or by low volume
samples.
For patient samples, the system
prompts you to analyze the
sample as a microsample. When
this happens, CO-ox results are
not reported and QC samples are
discarded and no results are
reported.
Message
Probable Cause
Corrective Action
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...