Rapidlab 1200 Operator’s Guide: Operating the System
2-35
02087462 Rev. V
3. Remove statistical outliers from the data by eliminating values outside
of ±3 SD.
NOTE:
Ensure that you include a significant number of results at the
extremes of the concentration ranges. If you do not, it could compromise the
quality of the correlation.
4. Perform a regression analysis.
Perform the analysis using a computer that can calculate the regression by the
Deming algorithm. This algorithm provides debiased slope and offset and is
the recommended method. A less acceptable alternative is to perform the
linear regression on a calculator or computer that does not use the Deming
algorithm.
The correction equation must be in the following form:
y = mx + b
where m is the slope and b is the offset
Use the results from the Rapidlab 1200 systems as the independent variable
(x). Use the results from the reference analyzer as the dependent variable (y).
NOTE:
You must perform the calculations as described above. The
calculation of a correction equation is the mathematical inverse of the
traditional correlation equation. Failure to calculate the regression as
described above produces an inaccurate correlation with the reference
analyzer.
5. Enter the slope and offset values.
Refer to
Combining Results for an a-v Study Report
Use this procedure to combine the results from an arterial and a mixed-venous
blood sample to create an a-v study report. You can combine results only for
Rapidlab 1245 or 1265 systems. Arterial blood sample results can only be
combined with mixed-venous blood sample results and not venous blood sample
results.
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...