6-32
Rapidlab 1200 Operator’s Guide: Troubleshooting
02087462 Rev. V
b. Rotate the gray wheel towards you, pull the tubing through the fluid detector, and
remove the tubing from the steel measurement module connector.
c. Push the new tubing through the fluid detector.
d. Connect the tubing to the steel measurement module connector and to the sample
chamber connector.
Use a clamp or tweezers to help you attach the tubing.
3. Remove obstructions from the CO-ox waste tubing:
a. Disconnect the tubing from the steel fluid connector and the connector in the
pump tubing.
Figure 6-16 CO-ox Sample Path
b. Push a clot-removal line through the tubing, opposite the direction of sample flow.
c. Reconnect the CO-ox waste tubing.
4. Remove obstructions from the waste path in the reagent manifold:
a. Disconnect the CO-ox pump tubing from the reagent manifold tubing at the
fitting.
1
Steel fluid connector
Summary of Contents for Rapidlab 1200
Page 1: ...Rapidlab 1200 Systems Operator s Guide 02087462 Rev V 2010 01 ...
Page 6: ...iv Rapidlab 1200 Operator s Guide Using This Guide 02087462 Rev V ...
Page 88: ...1 68 Rapidlab 1200 Operator s Guide System Overview and Intended Use 02087462 Rev V ...
Page 126: ...2 38 Rapidlab 1200 Operator s Guide Operating the System 02087462 Rev V ...
Page 206: ...5 64 Rapidlab 1200 Operator s Guide Maintenance 02087462 Rev V ...
Page 290: ...6 84 Rapidlab 1200 Operator s Guide Troubleshooting 02087462 Rev V ...
Page 344: ...A 4 Rapidlab 1200 Operator s Guide Safety Instructions 02087462 Rev V ...
Page 354: ...B 10 Rapidlab 1200 Operator s Guide Service Ordering and Warranty 02087462 Rev V ...
Page 360: ...D 4 Rapidlab 1200 Operator s Guide Supplies 02087462 Rev V ...
Page 448: ...I 8 Rapidlab 1200 Operator s Guide Index 02087462 Rev V ...