Rapidlab 1200 Operator’s Guide: Maintenance
5-57
02087462 Rev. V
Flushing the Tubing for the AutomaticQC Module
1. Connect the syringe tubing to the AutomaticQC manifold.
2. Flush with a minimum of 3 mL water.
Figure 5-31 AutomaticQC Tubing
3. Flush with a minimum of 12 mL air.
4. Collect the water in an absorbent swab or equivalent.
5. Disconnect the syringe tubing.
1
Water collection area
2
AutomaticQC manifold