Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
7-13
Input Failure
Set whenever there is a communication error between the Fieldbus card
and the OCX.
Simulation Active
Set whenever the Fieldbus Simulate Switch is set to ON at the Fieldbus
card or software simulate option enabled.
Other Error
Set whenever XD_ERROR is non-zero.
TRANSDUCER BLOCK
The Transducer Block was designed to provide the information necessary to
interface OCX 8800 to the Fieldbus.
Transducer Block
Parameters
Table 7-3 gives a description of all parameters, or gives the location of the
Fieldbus specifications the description can be found.
Table 7-3. Transducer Block Parameter Description
Parameter Mnemonic
Valid Range
Units
Description
ALARM_RELAY_EVENT1
See Table 7-7
Enumerated
The first of three conditions that cause the alarm output to turn on..
ALARM_RELAY_EVENT2
See Table 7-7
Enumerated
The second of three conditions that cause the alarm output to turn on.
ALARM_RELAY_EVENT3
See Table 7-7
Enumerated
The third of three conditions that cause the alarm output to turn on.
ALARM_RELAY_STATE
0: Off
– 1: On
Enumerated
The state of the alarm output.
ALERT_KEY
See FF-891 section 5.3.
ANALYZER_SW_BUILD_
The date the analyzer software was built.
DATE
ANALYZER_SW_BUILD_
The build number of the analyzer software.
NUMBER
ANALYZER_SW_
The checksum of the analyzer software.
CHECKSUM
ANALYZER_SW_VERSION
The version of the analyzer software.
BLOCK_ALM
See FF-891 section 5.3.
BLOCK_ERR
See FF-891 section 5.3.
BLOWBACK_DURATION
1-5
Seconds
The amount of time the blowback solenoid will be on.
BLOWBACK_ENABLED
0: No
– 1: Yes
Enumerated
Enables or disables the automatic blowback cycle.
BLOWBACK_INTERVAL
0-32767
Minutes
The time between blowback cycles.
BLOWBACK_PURGE_TIME
0-500
Seconds
The amount of time before returning the output to process after performing
a blowback.
BLOWBACK_STATE
0, 1, 2
Enumerated
The current state of the blowback cycle. (0=Idle, 1=Blow, 2=Purge)
CAL_GAS_TIME
60-1200
Seconds
The amount of time calibration gas should flow before a reading is taken.
CAL_PURGE_TIME
60-1200
Seconds
The amount of time before returning the output to process after calibrating.
CAL_REC_ENABLE
0: No
– 1: Yes
Enable/disable calibration recommended alarm.
CAL_RESULTS
See Table 7-10
Bit Enum
Calibration result.
CAL_STATE
See Table 7-4
Enumerated
The current state of the calibration cycle.
CAL_STATE_STEP
Initiates a calibration or goes to the next calibration step.
CAL_STATE_TIME
Seconds
The time left in the current calibration step.
COLLECTION_DIRECTORY
See Transducer Block Specification, part 1. FF-902, page 11.
COMB _SENSOR_CAL_LOC
See FF-903 section 3.3.
COMB_SENSOR_CAL_
See FF-903 sections 3.3 and 4.5.
METHOD
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...