Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
9-18
Figure 9-14. O
2
Cell, Heater,
and Thermocouple, Exploded
View
Disassemble O
2
Cell and Heater Strut Assembly
4
Do not attempt to replace the O
2
cell until all other possibilities for poor
performance have been considered. If cell replacement is needed, order the
O
2
cell replacement kit (Refer to Section 10, Replacement Parts).
6
The O
2
cell replacement kit contains an O
2
cell and flange assembly, gaskets,
7
socket head cap screws, and anti-seize compound. The items are carefully
8
packaged to preserve precise surface finishes.
9
Do not remove items from the package until they are ready to be used.
12
1. Remove the four allen cap screws (1, Figure 9-14) from the O
2
cell (2).
Remove the O
2
cell. The cell flange has a notch that may be used to
11
gently pry the flange away from heater tube (3).
10
1.
Screw
2.
O
2
Cell
3.
Heater Tube
4.
Contact/
Thermocouple
Assembly
5.
Gasket
6.
Screw
7.
Lockwasher
8.
Return Wire
9.
Heater Strut
Assembly
10.
Spring Clip
11.
Spring
12.
Strut Bracket
3
5
Test Gas
Passage
Holes
2
1
NOTE
The pad on the end of contact/thermocouple assembly (4) will sometimes
fuse to the O
2
cell (2).
2. If the O
2
cell is fused to the contact pad, push the O
2
cell back into the
heater tube (against spring pressure) and quickly twist the O
2
cell. The
cell and contact pad should separate. If the contact pad stays fused to
the cell, a new contact/thermocouple assembly (4) must be installed.
3. Remove and discard gasket (5). Clean the mating surface of heater
tube (3). Remove burrs and raised surfaces with a block of wood and
crocus cloth.
4. Remove screws (6), lockwashers (7), return wire (8), and heater strut
assembly (9).
5. If replacing contact and thermocouple assembly (4), use a pencil to
mark location of spring clip (10) before removing. Squeeze tabs on
spring clip to remove. Retain spring clip and spring (11); replace if
damaged.
6. While carefully handling new contact and thermocouple assembly (4)
lay old assembly next to new one. Transfer match marks to new
assembly.
7. Carefully guide new contact and thermocouple assembly (4) through
strut bracket (12), spring (11), and spring clip (10) until spring clip
reaches pencil mark.
Use care when handling contact and thermocouple assembly. The ceramic rod in this
assembly is fragile.
Do not remove the O
2
cell unless you are certain it needs to be replaced. Removal may
damage the cell and platinum pad. Remove the O
2
cell for cell replacement only.
3
7
3
9
0
0
32
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...