Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
5-14
D/A TRIM
PROCEDURES - HART
O
2
D/A trim procedure using HART
Use the following procedure to perform the O
2
D/A trim procedure using the
Field Communicator. If necessary, refer to Section 6, Field Communicator, for
the HART menu tree.
NOTE
To select a menu item, either use the up and down arrow keys to scroll to the
menu item and press the right arrow key or use the number keypad to select
the menu item number.
To return to a preceding menu, press the left arrow key.
1. From the
DIAG/SERVICE
menu, select
D/A TRIM
. Select
O2 D/A Trim
.
2. Press the right arrow key to start the procedure. (If you wish to exit D/A
Trim with no changes, select
ABORT
.)
3. The Field Communicator displays
WARNING: Loop should be
removed from automatic control
. Remove the OCX 8800 from any
automatic control loops to avoid a potentially dangerous operating
condition and press
OK
.
4. The Field Communicator displays
Connect reference meter to O2
output
.
5. Remove the electronics housing cover.
6. Refer to Figure 2-6. Connect a digital multimeter to read the milliamp
output from the O
2
D/A converter circuit. Connect the positive lead to
the AOUT1+ terminal and connect the negative lead to the AOUT1-
terminal. Then, press
OK
at the Field Communicator.
7. The Field Communicator displays
Setting Fld dev output to 4 mA
.
Press
OK
. Read the O
2
millamp output at the digital multimeter. Enter
the reading at the Field Communicator and press
ENTER
. (Select
ABORT
to exit without changes).
8. The Field Communicator displays
Setting Fld dev output to 20 mA
.
Press
OK
. Read the O
2
millamp output at the digital multimeter. Enter
the reading at the Field Communicator and press
ENTER
. (Select
ABORT
to exit without changes).
9. The Field Communicator displays
Setting Fld dev output to 4 mA
.
Press
OK
.
10. The Field Communicator displays
Fld dev output 4.00 mA equal to
reference meter?
Using the up or down arrow, select
1 Yes
or
2 No
and Press
ENTER
. If
No
, the process repeats from step 6.
11. The Field Communicator displays
Setting Fld dev output to 20 mA
.
Press
OK
.
12. The Field Communicator displays
Fld dev output 20.00 mA equal to
reference meter?
Using the up or down arrow, select
1 Yes
or
2 No
and Press
ENTER
. If
No
, the process repeats from step 7.
13. The Field Communicator displays
NOTE: Loop may be returned to
automatic control
.
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...