Instruction Manual
IM-106-880, Rev 2.1
April 2017
OCX 8800
9-17
Figure 9-13. Removal of Sample
and Exhaust Tubes
Remove Sample and Exhaust Tubes
1. Secure the sensor housing (1, Figure 9-13) in soft (plastic, wood, or
brass) vice jaws.
2. Use a propane torch to heat the sample tube (2) or exhaust tube (3) to
450°F (232°C), minimum. Apply the heat near the threaded end of the
tube.
3. While heating the tube, use a pipe wrench to apply removal torque to
part being removed. Apply torque until the pipe thread sealant softens.
Remove and discard the used sample tube (2), exhaust tube (3), or
in-situ filter (4).
4. Use MEK or methylene chloride solvent to clean thread sealant residue
from the internal pipe threads in the housing. Refer to applicable MSDS
sheet for solvent handling precautions.
1.
Sensor Housing
2.
Sample Tube
3.
Exhaust Tube
4.
In Situ Filter
4
Use heat resistant gloves when removing the probe tube or exhaust tube. The tubes are
bonded with a thread sealing compound. The compound softens at 450°F (232°C). The
heated parts can cause severe burns.
1
2
3
3
7
3
9
0
0
48
Summary of Contents for OCX 8800
Page 2: ......
Page 4: ......
Page 10: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 ii ...
Page 72: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 4 8 ...
Page 88: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 5 16 ...
Page 100: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 6 12 ...
Page 164: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 7 64 ...
Page 176: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 8 12 ...
Page 213: ......
Page 223: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 10 11 ...
Page 225: ......
Page 262: ...Instruction Manual IM 106 880 Rev 2 1 April 2017 OCX 8800 C 2 ...